Saline (IV drip) level monitor with nurse alert
In crowded Indian hospital wards one nurse watches many patients, and an empty IV bottle lets air and backflow into the line. A clip-on monitor that measures the drip rate and remaining fluid and alerts the nurse station prevents this.
- Parts
- ₹2,840
- Build
- 12 weeks
- Board
- ESP32
- Objectives
- 6
Block diagram
- IR drop sensor on the drip chamberCounts drops
- Load cellWeighs the remaining fluid
- ESP32Works out rate and time left
- Nurse station pageShows every bed
- Buzzer and LEDWarn at the bed
- 4G modemSends SMS alerts and calls over the mobile network+
- Cloud dashboardStores readings, plots trends and sends alerts+
Objectives
- To count drops and compute the drip rate in drops per minute and mL per hour.
- To measure remaining fluid by weight within ±10 mL.
- To alert the nurse station when 10% is left or the flow stops.
- To test against a timed manual drop count at three rates.
- To send SMS alerts through a 4G modem where no Wi-Fi is available.
- To publish readings over MQTT and show them on a live dashboard with threshold alerts.
Parts and specifications
- ESP32 DevKit V1Dual-core 240 MHz, 520 KB SRAM, Wi-Fi + Bluetooth, 3.3 V logic, 12-bit ADC (use ADC1 pins when Wi-Fi is on)₹450Buy
- IR photodiode (black)Daylight-blocking package, reverse-biased in a divider for beam-break sensing₹10Buy
- 3D-printed drip chamber clip and hook₹200
- IV set and 500 mL saline bottle (for testing with water)₹120
- A7670C 4G LTE module4G Cat-1 with 2G fallback, SMS, calls and data, works with any Indian SIM including Jio; 3.4 to 4.2 V, 2 A peaks₹1,500Buy
- LM2596 buck converter (set to 4.0 V for the modem)4 to 40 V in, 1.25 to 37 V out, up to 2 A continuous (3 A peak), about 85% efficient₹90Buy
Total about ₹2,840 at typical online prices. Tap a part for its pinout and datasheet notes.
Design calculations
Standard sets give 20 drops per mL. Flow (mL/h) = drops per minute × 60 / 20 = 3 × dpm. 30 dpm = 90 mL/h.
Remaining volume from weight: 1 g ≈ 1 mL for saline. 120 mL left at 90 mL/h = 80 minutes.
At 60 dpm drops are 1 s apart; a drop shadow lasts about 10 to 20 ms, so the interrupt catches each one.
No drop for 3× the expected interval (for example 6 s at 30 dpm) = blocked or empty.
Pin map
| Signal | Pin or connection |
|---|---|
| IR LED (through 100 Ω) | 3.3 V, always on |
| Photodiode → LM393 comparator out | GPIO 27 (interrupt) |
| HX711 DT / SCK | GPIO 16 / 4 |
| Buzzer / OLED | GPIO 26 / I²C 21, 22 |
Firmware outline
- Tare the hook with an empty bottle weight entered once.
- Count drops by interrupt; compute dpm from the last 10 intervals.
- Read weight every 5 s; convert to mL left.
- Alert at 10% left, flow stopped, or rate 30% away from the prescribed rate.
- Send bed number, rate, mL left and time left to the nurse station page every 10 s.
Test plan
| Test | Expected result |
|---|---|
| Manual count with stopwatch at 20, 40, 60 dpm | Within ±1 dpm |
| Weight against measuring cylinder | Within ±10 mL |
| Clamp closed | Flow-stopped alert within 10 s |
| Bottle near empty | 10% alert |
These rows start your results table when you plan the project.
Viva practice
Say your answer out loud first, then check. The first 2 answers are free; Maker opens all 5.
Make it yours
One or two upgrades for a mini project, two to four for a major one.
Build these first
Datasheets and sources to cite
- Avia Semiconductor HX711 datasheet
- Vishay TSAL6200 IR LED datasheet
- ISO 8536-4 (infusion sets: 20 drops ≈ 1 mL)
About this project
How much does the saline (iv drip) level monitor with nurse alert project cost?
About ₹1,250 for the parts in the base list: ESP32 DevKit V1, IR LED 940 nm, IR photodiode (black), LM393 dual comparator, Load cell 1 kg, HX711 load cell amplifier, Active buzzer 5 V, 0.96 inch OLED (SSD1306, I²C), 3D-printed drip chamber clip and hook, IV set and 500 mL saline bottle (for testing with water).
How long does it take?
About 8 weeks for the base project, built on an ESP32. Each upgrade adds time.
Which branches is it suitable for?
Electronics and Communication, Instrumentation, Biomedical, Computer Science / IT (IoT, AI), as a mini or major project.
Which pins and design calculations does it use?
IR LED (through 100 Ω) on 3.3 V, always on; Photodiode → LM393 comparator out on GPIO 27 (interrupt); HX711 DT / SCK on GPIO 16 / 4. Worked calculations on this page: drip rate, time left, drop timing, flow stopped.
How do I test it?
Manual count with stopwatch at 20, 40, 60 dpm: Within ±1 dpm. Weight against measuring cylinder: Within ±10 mL. Clamp closed: Flow-stopped alert within 10 s. Bottle near empty: 10% alert.
What are the objectives?
To count drops and compute the drip rate in drops per minute and mL per hour. To measure remaining fluid by weight within ±10 mL. To alert the nurse station when 10% is left or the flow stops. To test against a timed manual drop count at three rates.