Projects / Ice and boiling point calibration

Ice and boiling point calibration

Beginner1 hGuided buildReviewed

Build it step by step

0 of 7 done
  1. NANODS18B20NTC1027.4°C

    Wire both probes

    DS18B20 with a pull-up, NTC with the divider.

  2. NANO

    Read at room temperature

    Compare with each other.

  3. NANO

    Make an ice slurry

    Crushed ice with a little water, stirred.

  4. NANO

    Take the ice reading

    Type i.

  5. NANO

    Take the boiling reading

    Type b in a rolling boil.

  6. NANO

    Apply the correction

    Type r and read corrected values.

  7. NANO

    Check in warm water

    Compare with a reference thermometer.

In short

How do you build an ice and boiling point calibration?

Check temperature sensors at 0 °C in ice water and at boiling, and work out a correction. You need Arduino Nano, DS18B20 waterproof probe, NTC 10 kΩ waterproof probe, Resistor 4.7 kΩ and Resistor 10 kΩ. It takes 7 steps and about 60 minutes.

Level
Beginner
Time
About 60 min
Parts
5

Parts list

Tick what you already own. It saves to My Garage, so “Build with what I have” knows.

Wiring

FromToWhy
DS18B20 red and blackNano 5V and GNDPower
DS18B20 yellow (data)Nano D2, with 4.7 kΩ to 5VOne-wire data
NTC probe one leadNano 5VTop of the divider
NTC probe other leadNano A0 and a 10 kΩ resistor to GNDDivider output
USBNanoPower and serial monitor

Code

Two fixed points give a straight-line correction. Crushed ice and water, stirred, sits at 0 °C; boiling water sits at about 100 °C at sea level and a little lower higher up, by roughly 1 °C for every 300 m, which the code uses as the altitude correction. Type i with the probes in the ice slurry (mostly ice, just enough water to touch the probe) and b in a rolling boil, then r to see corrected values. The probes must not touch the bottom or sides of the container, which are hotter or colder than the water. Boiling water causes burns: use a steady pot on a stable surface, keep leads away from the burner, let it cool before handling, and keep the Arduino well away from steam. Check that your NTC probe is rated for 100 °C before dipping it in boiling water, as some cheap ones are rated lower. Use distilled water for the best results. Arduino IDE, board "Arduino Nano" (try the Old Bootloader option if the upload fails).

Runs on a virtual Arduino Nano in your browser, with sliders for the sensors. Happy with it? Verify and upload to your real board from the same page (Chrome or Edge on a computer).

// Two-point calibration at the ice point and the boiling point. Libraries: OneWire, DallasTemperature. Board: Arduino Nano.
// Serial monitor at 115200 baud. Type i in the ice slurry, then b in boiling water. Then r shows corrected readings.
#include <OneWire.h>
#include <DallasTemperature.h>

OneWire ow(2); DallasTemperature ds(&ow);
const float ALTITUDE_M = 920.0;                       // your height above sea level, for the boiling point
const float BETA = 3950.0, R0 = 10000.0, RFIX = 10000.0;
float iceDS = 0, boilDS = 100, iceNTC = 0, boilNTC = 100;
float refBoil;

float ntcC() {                                        // average 50 readings
  long s = 0; for (int i = 0; i < 50; i++) { s += analogRead(A0); delay(4); }
  float adc = s / 50.0;
  float r = RFIX * adc / (1023.0 - adc);              // NTC on top of the divider
  float t = 1.0 / (log(r / R0) / BETA + 1.0 / 298.15) - 273.15;
  return t;
}
float dsC() { ds.requestTemperatures(); return ds.getTempCByIndex(0); }

float avg(float (*f)(), int n) { float s = 0; for (int i = 0; i < n; i++) { s += f(); delay(200); } return s / n; }

void setup() {
  Serial.begin(115200); ds.begin();
  refBoil = 100.0 - ALTITUDE_M / 300.0;               // water boils about 1 C lower for every 300 m of height
  Serial.print("Reference boiling point here: "); Serial.println(refBoil, 2);
  Serial.println("Commands: i = ice reading, b = boiling reading, r = corrected readings");
}

void loop() {
  if (Serial.available()) {
    char c = Serial.read();
    if (c == 'i') { iceDS = avg(dsC, 20); iceNTC = avg(ntcC, 20); Serial.print("Ice DS18B20 "); Serial.print(iceDS, 2); Serial.print("  NTC "); Serial.println(iceNTC, 2); }
    if (c == 'b') { boilDS = avg(dsC, 20); boilNTC = avg(ntcC, 20); Serial.print("Boil DS18B20 "); Serial.print(boilDS, 2); Serial.print("  NTC "); Serial.println(boilNTC, 2); }
    if (c == 'r') {
      float gainDS = (refBoil - 0.0) / (boilDS - iceDS), gainNTC = (refBoil - 0.0) / (boilNTC - iceNTC);
      for (int k = 0; k < 10; k++) {
        float d = dsC(), n = ntcC();
        Serial.print("DS18B20 "); Serial.print((d - iceDS) * gainDS, 2); Serial.print("   NTC "); Serial.println((n - iceNTC) * gainNTC, 2);
        delay(500);
      }
    }
  }
}

Test checklist

What you should see at each check. If a reading is far off, stop and use the troubleshooting list.

CheckExpected
Ice slurryCorrected values within 0.2 °C of zero
Boiling waterWithin 0.5 °C of the local boiling point
Warm waterBoth probes agree within 0.5 °C
Room airBoth stay close to each other

If it does not work

DS18B20 reads -127

Check the 4.7 kΩ pull-up and the wiring.

NTC reads odd values

Check which side the 10 kΩ resistor is on in the divider.

Ice reading not zero

Stir the slurry and keep the probe off the walls.

Boiling reading far off

Check the altitude constant and that the water is truly boiling.

Still stuck? Find the fault step by step with your multimeter →

Ask Lark AI

Common questions

What parts do I need for the ice and boiling point calibration?

Arduino Nano, DS18B20 waterproof probe, NTC 10 kΩ waterproof probe, Resistor 4.7 kΩ and Resistor 10 kΩ.

How long does the ice and boiling point calibration take to build?

About 60 minutes for a beginner maker, in 7 steps.

How do I know the ice and boiling point calibration works?

Ice slurry: Corrected values within 0.2 °C of zero. Boiling water: Within 0.5 °C of the local boiling point. Warm water: Both probes agree within 0.5 °C.

DS18B20 reads -127: what should I check?

Check the 4.7 kΩ pull-up and the wiring.

NTC reads odd values: what should I check?

Check which side the 10 kΩ resistor is on in the divider.

Ice reading not zero: what should I check?

Stir the slurry and keep the probe off the walls.

Boiling reading far off: what should I check?

Check the altitude constant and that the water is truly boiling.